Problem
Poor performance
Possible Cause
The insert is not correctly
tightened on the handpiece
Insert broken, worn or
deformed
Insufficient or excessive
powder level in the container.
TROUBLESHOOTING
Solution
Unscrew the insert and screw
it back in correctly using the
Mectron torque wrench (See
Chapter 5.4 on page 32
Replace the insert with a
new one
Restore the correct level of
powder in the container (See
Chapter 4 on page 17).
)
115