Extract the peristaltic pump from its seat,
pulling it towards yourself. Use caution as
pieces may detach;
Attach the new peristaltic pump onto the
base until you hear it click in, and connect
the two pump tubes to the respective
couplings underneath it;
WARNING: Only use original spare
parts.
Refit the plastic protection onto the
peristaltic pump.