MAINTENANCE

If the appliance is not used for a long time
observe the following recommendations:
1.
Empty the powder containers;
2.
Run a complete cleaning cycle of
the irrigation circuit using the "flush"
function (see
Chapter 6 on page 40
for both the Ultrasound and Polishing
parts;
3.
Drain the irrigation circuits of any
residual water, removing the tank
and running the "flush" cycle for the
Ultrasound part;
4.
Eliminate the condensate water from
the air filter (see
Chapter 9.3 on page
101
);
5.
Disconnect the appliance from the
power mains and from the air and water
circuits;
9.1
Replacing the Peristaltic Pump
There is a plastic protection on the left
side of the appliance, which covers the
peristaltic pump housing. Remove this
protection by pressing on the sides and
pulling towards yourself;
CAUTION: Before carrying out
operations on the peristaltic pump,
make sure the appliance is unplugged
and that the liquids container is not
connected.
Remove the two tubes of the pump from
the respective clutches positioned below
it;
98
6.
Clean and dry the water filter (see
Chapter 9.2 on page 100
7.
In case of long periods of non-use,
place the appliance back in its original
packaging, in a safe place;
)
8.
Before using the appliance again,
clean and sterilise the handpiece and
accessories following the instructions in
Chapter 8 on page 49
9.
Check that the inserts are not worn,
deformed or broken, with particular
attention to the integrity of the tip.
WARNING: Periodically check the
integrity of the power supply cable; if
damaged, replace it with an original Mectron
spare part.