Fill the irrigation solution tank with 300ml
of distilled water;
NOTE: Use of distilled water is
recommended for long periods of
non-use. If the device is used after the
disinfection procedure, use common
running water.
Holding the irrigation solution tank in a
vertical position, insert it on the machine
core unit of the device until it comes into
contact;
CAUTION: Do not overturn the tank,
the cap is not watertight. The outflow
of physiological solution or aggressive
liquids may cause damage to the
surfaces.
Run a number of "flush" cycles (see points
7 to 10) sufficient to completely drain the
water from the tank;
After cleaning the irrigation circuit,
disassemble the individual parts (see
Chapter 7 on page 45
clean and sterilise them (see
page 49
).
CAUTION: If the handpiece is stowed
back in its support, with the "flush"
function selected
(flashing LED) but not running, the
function is deactivated;
CAUTION: During the "flush" cycle,
i.e. when the function has been selected
and activated by pressing the pedal,
if the handpiece is stowed back in the
support, the cycle is not interrupted.
CLEANING AND STERILISATION
19
20
21
22
) and proceed to
Chapter 8 on
73