Section 10 Maintenance And Service; Cleaning The Analyzer - Abaxis VETSCAN VS2 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Section 10
Maintenance and
Service
The VetScan VS2 Analyzer requires minimal maintenance.
Clean the exterior of the analyzer weekly with mild detergent
and a soft, damp cloth. The air filter requires cleaning once a
month. Regular maintenance of the analyzer will assure reliable
operation.
CAUTION:
Abaxis recommends only the cleaning methods
described in this section. If for some reason
another method is needed, contact Abaxis
Technical Support beforehand at
1-800-822-2947 to verify that the proposed
method will not damage the analyzer. Abaxis is
not responsible for damage caused by non-rec-
ommended cleaning methods.

10.1 Cleaning the Analyzer

Clean the analyzer's external case and display at least weekly, as
described below.
10.1.1 Cleaning the Case
Clean the analyzer with a soft cloth, dampened with a mild, non-
abrasive detergent or cleaning solution, such as Simple Green
a 10% bleach solution, or a 30% isopropyl alcohol solution. Pre-
soaked towelettes (isopropyl alcohol) may be used as an alter-
nate. Purchase isopropyl towelettes from computer retailers.
Do not spray or pour any detergents, solutions or other liquids
directly onto the analyzer. Dampen a soft cloth or disposable
paper towel with the detergent, then apply to the analyzer.
Note:
While cleaning, inspect the analyzer's casing to
make sure it is not cracked or damaged.
Maintenance and Service
®
,
10-1

Publicidad

Tabla de contenido

Solución de problemas

loading

Productos relacionados para Abaxis VETSCAN VS2

Tabla de contenido