Important; Automatic Ph Calibration - XS Instruments Revio Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Flashing of the red led every 3 seconds.
Acoustic signal (see paragraph Instrument Setup Menu /Buzzer Mode).
Temperature calibration
All the instruments of this series are pre-calibrated for a correct temperature reading. However, if there is a
difference between the measured and the real temperature (usually due to a probe malfunction), it is
possible to perform an offset adjustment of + 5°C.
After connecting the temperature probe in the correct measurement channel, use the keys
to correct the temperature offset value and confirm with
Restore Factory Default
If the instrument does not work properly or incorrect calibrations have been carried out, confirm Yes with
the button
to take all the parameters of the pH menu back to the default settings.
IMPORTANT: the factory reset of the parameters does not delete the stored data.

• Automatic pH calibration

Example: three-point calibration with USA type buffers
In pH measure mode keep the button
Rinse the electrode with distilled water
Press the button
beaker on display). The first calibration point is always the neutral pH (7.00 for USA curve, 6.86 for NIST
and DIN curve), while the rest are at user's discretion.
When the icon
appears, confirm the first point by pressing the button
The measured value flashes on the display and then the icon of the pH 7.00 beaker
bottom left, indicating that the instrument is calibrated on the neutral point.
Remove the electrode, rinse with distilled water
Press the button
solution. In the beaker the different pH-buffers, that the device can recognize automatically, scroll.
When the 4.01 value is recognized and the icon
The actual measured value flashes on the display and subsequently, next to the beaker pH 7.00, the
icon of the beaker pH 4.01
UM ReVio EN rev.0 04.06.2020
pressed for 3 seconds to enter calibration mode.
and gently dab with paper towel.
and dip the electrode in the pH 7.00 buffer solution (as indicated by the
in order to proceed with the calibration and dip the sensor in the pH 4.01 buffer
4.01
appears, indicating that the instrument is calibrated in the acid field.
19
.
and dab gently with absorbent paper.
appears, confirm by pressing the button
and
.
7.00
appears at the
.
EN