Tabla de contenido

Publicidad

EMC tables

Guidance and MANUFACTURER'S declaration – ELECTROMAGNETIC EMISSIONS
– FOR ALL THE EQUIPMENT AND THE SYSTEMS
I-PRESSS 4 is intended for use in the electromagnetic environment specified below. The
customer or the user of the I-PRESS 4 should assure that it is used in such an
environment.
Emission test
RF emissions
Cispr 11
RF emissions
Cispr 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and MANUFACTURER'S declaration – ELECTROMAGNETIC IMMUNITY
– FOR ALL THE EQUIPMENT AND THE SYSTEMS
I-PRESS 4 is intended for use in the electromagnetic environment specified below. The
customer or the user of I-PRESS 4 should assure that is used in such environment.
Immunity test
Electrostatic
discharge (ESD)
EN 61000-4-2
IACER SRL
Compliance
Group 1
Class B
Class A
Complies
Complies
Test
level
EN
60601-1-2
± 6kV contact
± 8kV air
Pagína 23
Electromagnetic environment –
guidance
I-PRESS 4 uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment
I-PRESS 4 is suitable for use in all
establishments,
establishments
connected to the public low-voltage
power supply
buildings used for domestic purposes.
The I-PRESS 4 is suitable for use in all
establishments, other than domestic
establishments
connected to the public low voltage
power supply network that supplies
buildings used for domestic purposes
Compliance level
± 6kV contact
± 8kV air
including
domestic
and
those
directly
network that supplies
and
those
directly
Electromagnetic
environment
guidance
Floors
should
be
wood, concrete or
ceramic tile. If floor
are
covered
with
synthetic
material,
the relative humidity
should be at least
30%
MNPG222-00
24/03/16

Publicidad

Tabla de contenido
loading

Tabla de contenido