• The smartLAB
it, wash your clothes with soap and water.
• Close the bottle tightly after use.
• Left over control solution should not be added back into the
control bottle.
• Store the bottle of smartLAB
temperature, below 30°C (86°F). Do not freeze nor refrigerate.
Running a smartLAB
Please make sure you have all necessary items at hand when run-
ning a control solution test. You need the smartLAB
smartLAB
pro test strip, and smartLAB
®
1. Switch on the meter
Take a strip out of the vial and immediately close it. Insert the strip
into the meter in the direction of the arrow on the strip. The meter
turns on automatically. „Code 888" appears on the display and af-
terwards the symbol for applying blood.
In case „Code 888", DOES NOT appear on the display,
please contact your supplier (see chapter „Note on
smartLAB
2. Running the control solution test
When the symbol for applying blood appears, shortly shake the bot-
tle with the smartLAB
the tip of the bottle with a tissue. Squeeze a little drop of control
solution on a clean, not adsorbing surface (plastic, glass, etc.) and
control solution can stain clothing. If you spill
®
control solution test
®
®
„NO CODE" test strips")
control solution. Open the bottle and wipe
®
control solution at room
®
control solution.
®
mini meter, a
®
24