Measurements

Measurement with the Desired Systolic Pressure

Model UA-767F is designed to detect the pulse and to inflate the cuff to a systolic pressure level automatically.
Use this method when re-inflation occurs repeatedly or when the results are not displayed even if the pressure
decreases to 20 mmHg or less.
1. Place the cuff on the arm at heart level (preferably the
left arm).
2. Press and hold the START button until a number about
30 to 40 mmHg higher than your expected systolic pressure
appears.
3. When the desired number is reached, release the START
button to start measurement. Continue to measure your
blood pressure as described on the previous page.
Notes for Proper Measurement
q Sit down in a comfortable position. Place your arm on a
table with your palm facing upward and the cuff at the same
level as your heart.
q Relax for about five to ten minutes before taking a
measurement. If you are excited or depressed by emotional
stress, the measurement will reflect this stress as a higher
(or lower) than normal blood pressure reading and the
pulse reading will usually be faster than normal.
q An individual's blood pressure varies constantly, depending on what you are doing and what you have eaten. What
you drink can have a very strong and rapid effect on your blood pressure.
q This device bases its measurements on the heartbeat. If you have a very weak or irregular heartbeat, the device
may have difficulty determining your blood pressure.
q Should the device detect a condition that is abnormal, it will stop the measurement and display an error symbol.
Refer to page 6 for the description of symbols.
q This blood pressure monitor is intended for use by adults only. Consult with your physician before using this device
on a child. A child should not use this device unattended.
E-12
At heart level
Press and hold
the button
Release the button
at the desired
systolic pressure
Refer to the previous
page for measurement
START