Table 4: Guidance and Manufacturer's Declaration - Electromagnetic Immunity
E
The ConMed Linvatec VP4726 and VP4726F Monitors are intended for use in the electromagnetic
environment specified below. The customer or the user of this VP4726 or VP4726F Monitor
Immunity Test
Conducted RF
IEC 61000-4-6
should assure that it is used in such an environment.
IEC 60601
Compliance Level
Test Level
3 Vrms
3 Vrms
150 kHz to 80 MHz
150 kHz to 80 MHz
E-44
Electromagnetic environment
guidance
Portable and mobile RF
communications equipment should
be no closer to any part of the
VP4726 and VP4726F Monitors
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended Separation
Distance
√ P
d
= [3.5 / V1]
,