4. Controlling
1. Switch knob 3 (set/measuring) to „set". Adjust the desired set point at knob 5 (set point adjustion).
2. Turn switch 3 back to „µS". The unit is now ready to use. As soon as the adjusted set point is exceeded, the
unit activates the socket at the main plug.
5. Calibration
If you change the measuring range between fresh and salt water, the unit has to be recalibrated.
For fresh water, a calibration solution of 250 µS and for salt water a solution of 50 mS is recommended. Immerse
the electrode in a calibration liquid of the corresponding range and the exact temperature. To adjust the display,
turn the calibration knob (6 or 7) of the corresponding range.
For fresh water, a second calibration fluid with 500 µS/cm resp. for salt water with 50 mS/cm is necessary. Turn
a small screw driver through the opening at the bottom of the unit until the display shows the right value. If the
unit shows a wrong value, it has to be recalibrated.
6. Electrode
The conductivity electrode has a plastic shaft. The contact pins are made from carbon and are corrosion safe,
also in sea water in long term use. The electrode is supplied with a thermosensor for automatic temperature
compensation. The electrode is maintenance-free.
Technical Data
Display:
Measuring range:
Resolution:
Accuracy:
Max. power on relay:
Working temperature:
Temperature compensation:
Humidity:
Control range:
Control accuracy:
Power supply:
Dimensions:
Weight:
7. Warranty
Should any defect in materials or workmanship be found within 24 months of the date of purchase, AB Aqua
Medic undertakes to repair, or at our option replace, the defective part free of charge – always provided the
product has been installed correctly, is used for the purpose that was intended by us, is used in accordance with
the operating instructions and is returned to us carriage paid. Proof of Purchase is required by presentation of
the original invoice or receipt indicating the dealer's name, the model number and date of purchase, or a
Guarantee Card if appropriate. This warranty may not apply if any model or production number has been altered,
deleted or removed, unauthorised persons or organisations have executed repairs, modifications or alterations,
or damage is caused by accident, misuse or neglect. We regret we are unable to accept any liability for any
consequential loss. Please note that the product is not defective under the terms of this Warranty where the
product, or any of its component parts, was not originally designed and / or manufactured for the market in
which it is used. These statements do not affect your statutory rights as a customer.
AB Aqua Medic GmbH - Gewerbepark 24 - 49143 Bissendorf/Germany
5
0.5 " LED, 3-1/2 digits
0 – 1,999 µS
0 - 199,9 mS
1 µS ( 0 – 1,999 µS)
0,1 mS ( 0 - 199,9 mS)
+/- 1% (+ 2 digits)
5 Amp. at 240 V (1,000 W)
0 - 50 °C
0 - 50 °C, automatic
below 80%
0 - 199 mS, digitally adjustable
+/- 5 µS, +/- 0,5 mS(+ 1 digit)
220 V, 50 Hz
150 x 85 x 40 mm
550 g
- Technical changes reserved – 1/2018