Guidance and manufacturer's declaration - electromagnetic immunity
The pump is intended for use in the electromagnetic environment specified below. The customer or
the user of the pump should assure that it is used in such an environment.
Immunity Test
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV line(s) to
line(s)
IEC 61000-4-5
±2 kV line(s) to
earth
<5% U
Voltage dips, short
(>95% dip in U
interruptions and
voltage variations
for 0.5 cycle
on power supply
input lines
40% U
(60% dip in U
IEC 61000-4-11
for 5 cycles
70% U
(30% dip in U
for 25 cycles
<5% U
(>95% dip in U
for 5 sec
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to the application of the test level.
T
IEC 60601 Test
Compliance
Level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
N/A
±1 kV line(s) to
line(s)
N/A
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
Electromagnetic environment -
Level
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical domestic commercial or
hospital environment.
Mains power quality should be that
of a typical domestic commercial or
hospital environment.
Mains power quality should be that
T
of a typical domestic commercial or
)
T
hospital environment.
If the user of the pump requires
T
continued operation during mains
)
T
power interruptions, it is
recommended that the pump be
powered from an uninterrupted
T
power supply or a battery.
)
T
T
)
T
Power frequency magnetic fields
3 A/m
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
32
guidance